Temperature Controlled Demountable Liquid Cells

Temperature Controlled Demountable Liquid Cells are intended for transmission measurements of liquids and mulls at temperatures ranging from 0°C to 240°C. These leak-free transmission cells are designed for both st ... read more

Temperature Controlled Demountable Liquid Cells
Add to Cart
$0.00

Info

Info

Parent Product:
Yes
Summary:
  • Pathlengths from 6µm to 1mm
  • Temperature range:  0°C to 240°C
  • Maximum pressure: 4.5 MPa

Temperature Controlled Demountable Liquid Cells are intended for transmission measurements of liquids and mulls at temperatures ranging from 0°C to 240°C. These leak-free transmission cells are designed for both static and flow operation. In addition, these 316 stainless steel transmission cells feature high chemical resistance and readily exchanged pathlength spacers. The Temperature Controlled Demountable Liquid Cells are valuable for infrared measurements of liquids, mulls and smears at elevated and reduced temperatures. They are useful in process control and laboratory applications.

Features

  • Variable pathlength, from 6µm to 1mm.
  • Heatable to 200°C withmFKM fluoroelastomer (equivalent to Viton) o-rings or 240°C with optional FFKM perfluoroelastomer (equivalent to Kalrez) o-rings.
  • Coolable to 0°C with suitable coolant, purge and o-rings.
  • Flow-through cooling conduit stabilizes the cell temperature.
  • Thermally isolated cell body prevents heat transfer to the spectrometer.
  • Leak-free design utilizes standard, non-drilled windows.
  • Easily disassembled for cleaning and window replacement.
  • Two ports permit static or flow applications.
  • Chemically resistant 316 stainless steel cell body with Luer Lok or 1/8" Swagelok™ fittings.
  • Sample slide plate mounted for easy installation in FTIR spectrometers.
  • Available with an 8mm or 20mm diameter open aperture.
  • Maximum operating pressure with ZnSe windows:
    • 4.5 MPa (660 psi) for the TFC-3-S13 (8 mm aperture).
    • 690 kPa (100 psi) for the TFC-3-S25 (20 mm aperture).
  • Temperature controlled cells for operation at lower or higher temperatures available on special order.

Includes

  • Stainless steel cell body with Luer Lok or Swagelok™ fittings.
  • Embedded K-type thermocouple, cartridge heaters and cooling conduit.
  • FKM fluoroelastomer (equivalent to Viton) o-rings.
  • Slide plate mounting.
  • One PTFE dozen spacers. One each for the following pathlengths: 6, 12, 25, 56, 100, 150, 250, 390, 500, 630, 750, and 950 µm.
  • Windows need to be ordered separately.